Section 3 – Installation
Page 13 09/11/2008 REV 0.99
3.0 INSTALLATION
This section provides information and procedures on installing and wiring the recorder. Included are handling
procedures, installation and wiring specifications, and instructions for both standard and optional equipment.
3.1 Equipment Handling
3.1.1 Initial Inspection
Exercise care when unpacking the instrument from the shipping carton. The instrument is packed in a
shockproof foam retainer to prevent damage during normal transit. If damage to the shipping carton is
evident, ask the carrier‟s representative to be present when the instrument is unpacked.
3.1.2 Unpacking Procedure
To unpack your recorder, first remove the foam retainer and instrument from the shipping carton. Then,
carefully remove the instrument from the foam retainer.
3.1.3 Detected Damage
If damage is detected after unpacking the instrument, re-pack the instrument and return it to the factory as
described in the following section.
3.1.4 Equipment Return
Before returning a damaged or malfunctioning instrument to the factory for repairs, contact the sales
organization from which you purchased the instrument. A Return Merchandise Authorization number must be
obtained from the factory before returning an instrument for any reason.
3.1.5 Storage
For prolonged storage before installation, re-pack the recorder in the shipping container. Cushion the
recorder with foam molding or an equivalent and store in a cool, dry area. It is not recommended to store the
recorder for more than one year. If longer storage time is required, contact the factory for additional storage
information.
3.2 Environment and Location
The instrument is intended to operate in the following environment:
Indoor Use Only
Installation Category II per IEC 664
Pollution Degree Level II per UL3111-1 / IEC61010-1
Altitude up to 2,000 m
Temperature -10 °C to +50 °C operating per UL3111-1 / IEC61010-1
Humidity Maximum relative humidity 80% for temperatures up to
31°C decreasing linearly to 50% relative humidity at 40°C
AC Mains Supply 100 - 240 Vac ~ 50/60 Hz 35 Vac ±10%