Anritsu Series MS278XB Network Router User Manual


 
Measurements Operation Verification
MS278XB OM 5-71
5-6 Operation Verification
This section provides procedures that you can use to verify the signal analyzer is operating properly and meets
some of its specifications. You may perform the following verifications:
Reference Oscillator Aging Rate (Optional)
Frequency Readout Accuracy
Frequency Span Accuracy
Swept Resolution Bandwidth
Single Sideband Phase Noise
Average Noise Level
Frequency Response
Reference Level Switching Uncertainty
Resolution Bandwidth (RBW) Switching Uncertainty
Residual Spurious Response
Input-related Spurious Response
Refer to the Required Equipment list before beginning these verification procedures.
Required Equipment
Table 5-2 lists the equipment used throughout the verification procedures.
Note: Specifications listed in this section are for reference only and should be verified with the Signature
Datasheet, part number: 11410-00396. The most current revision can be downloaded from the Documents
area of the Anritsu Internet site: http://www.us.anritsu.com
Table 5-2. Required Equipment
Instrument Critical Specification Recommended Manufacturer/Model
Frequency Standard Frequency: 10 MHz
Accuracy: 5
× 10
–12
parts/day
Symmetricom Model 9390-9600
Synthesized Signal
Generator
Frequency Range: 10 MHz to 8 GHz
Ultra Low Phase Noise
Anritsu Model MG3691B with
Options 2A, 3, 4, 16
Synthesized Signal
Generator
Frequency Range: 10 MHz to 6 GHz
Spurious Performance <–85 dBc
HP 8665B
Adapter N(m) to K(f) Anritsu Model 34NKF50
Adapter Ruggedized K(m) to N(f) Anritsu Model 34RKNF50
Attenuator K(m) to K(f)
Frequency Range: DC to 8 GHz
Attenuation: 3 dB
Anritsu Model 43KB-3
Attenuator N(m) to N(f)
Frequency Range: DC to 8 GHz
Attenuation: 3 dB
Weinschel Model 1-3
Attenuator N(m) to N(f)
Frequency Range: DC to 8 GHz
Attenuation: 10 dB
Weinschel Model 4410
Cable N(m) to N(m)
Frequency Range: DC to 8 GHz
Impedance: 50 Ω
Any