12. Repeat steps 9 through 12 for the other combinations of RBW and Span settings.
Table 5-6. Swept Resolution Bandwidth Test
Frequency RBW Span
Specification
Minimum Maximum
100 MHz 10 Hz 30 Hz 9.8 Hz 10.2 Hz
20 Hz 50 Hz 19.6 Hz 20.4 Hz
30 Hz 100 Hz 29.4 Hz 30.6 Hz
50 Hz 200 Hz 49 Hz 51 Hz
100 Hz 300 Hz 98 Hz 102 Hz
200 Hz 500 Hz 196 Hz 204 Hz
300 Hz 1 kHz 294 Hz 306 Hz
500 Hz 2 kHz 498 Hz 510 Hz
1kHz 3kHz 980Hz 1.02kHz
2kHz 5kHz 1.96kHz 2.04kHz
3kHz 10kHz 2.94kHz 3.06kHz
5 kHz 20 kHz 4.9 kHz 5.1 kHz
10 kHz 30 kHz 9.8 kHz 10.2 kHz
20 kHz 50 kHz 19.6 kHz 20.4 kHz
30 kHz 100 kHz 29.4 kHz 30.6 kHz
50 kHz 200 kHz 49 kHz 51 kHz
100 kHz 300 kHz 98 kHz 102 kHz
200kHz 500kHz 196kHz 204kHz
300kHz 1MHz 294kHz 306kHz
500kHz 2MHz 490kHz 510kHz
1MHz 3MHz 980kHz 1.02MHz
2MHz 5MHz 1.96MHz 2.04MHz
3 MHz 10 MHz 2.7 MHz 3.3 MHz
5 MHz 20 MHz 4.5 MHz 5.5 MHz