Calibrating for Increased Measurement Accuracy
Frequency Response Error Corrections
Frequency Response Error Corrections
You can remove the frequency response of the test setup for the following measurements:
reflection measurements
transmission measurements
combined reflection and transmission measurements
Response Error Correction for Reflection Measurements
1. Press .
2. Select the type of measurement you want to make.
If you want to make a reflection measurement on PORT 1 (in the forward direction,
S
11
), leave the instrument default setting.
For ES analyzers, if you want to make a reflection measurement on PORT 2 (in the
reverse direction, S
22
), press:
3. Set any other measurement parameters that you want for the device measurement:
power, sweep type, number of points, or IF bandwidth.
4. To access the measurement error-correction menus, press:
5. If your calibration kit is different than the kit specified under the
softkey, press:
(select your type of kit)
If your type of calibration kit is not listed in the displayed menu, refer to "Modifying
Calibration Kits" on page 7-57.
6. To select a response correction, press:
Connect the short or open calibration standard to the port you selected for the test port
(PORT 1 for S
11
or PORT 2 for S
22
on ES analyzers).
NOTE Include any adapters or cables that you will have in the device measurement.
That is, connect the standard device to the particular connector where you
will connect your DUT.
Preset
Meas
Refl: REV S22 (B/R)
Cal
CALIBRATE MENU
Cal
CAL KIT [ ]
Cal
CAL KIT
SELECT CAL KIT
RETURN
Cal
CALIBRATE MENU
RESPONSE