6-51
Calibrating for Increased Measurement Accuracy
Calibrating for Non-Coaxial Devices (ES Analyzers Only)
6. For the purposes of this example, change the name of the standard by pressing:
and modifying the name to "LINE."
7. When the title area shows the new label, press:
Assign the Standards to the Various TRL Classes
8. To assign the calibration standards to the various TRL calibration classes, press:
9. Since you previously designated standard #1 for the REFLECT standard, press:
10.Since you previously designated standard #6 for the LINE/MATCH standard, press:
11.Since you previously designated standard #4 for the THRU/LINE standard, press:
12.To complete the specification of class assignments, press:
Label the Classes
NOTE To enter the following label titles, an external keyboard may be used for
convenience.
13.Press .
14.Change the label of the "TRL REFLECT" class to "TRLSHORT."
15.Change the label of the "TRL LINE OR MATCH" class to "TRLLINE."
16.Change the label of the "TRL THRU" class to "TRLTHRU."
17.Press .
Label the Calibration Kit
18.Press and create a label up to 8 characters long. For this example, enter
"TRL KIT1" DONE.
19.To save the newly defined kit into nonvolatile memory, press:
NOTE Refer to "Saving Modified Calibration Kits to a Disk" on page 7-65 for
information about saving modified calibration kits, along with calibration
data and instrument states, to a disk.
LABEL STD
DONE
STD DONE (DEFINED)
Cal
CAL KIT
MODIFY
SPECIFY CLASS
MORE
MORE
TRL REFLECT
1 x1
TRL LINE OR MATCH
6 x1
TRL THRU
4 x1
SPECIFY CLASS DONE
LABEL CLASS
MORE
MORE
LABEL CLASS DONE
LABEL KIT
KIT DONE (MODIFIED)
SAVE USER KIT