Calibrating for Increased Measurement Accuracy
Frequency Response and Isolation Error Corrections
5. If your calibration kit is different than the kit specified under the softkey,
press:
(select your type of kit)
If your type of calibration kit is not listed in the displayed menu, refer to "Modifying
Calibration Kits" on page 7-57.
6. To select a response and isolation correction and to start with the response portion of
the calibration, press:
7. Connect the short or open calibration standard to the port you selected for the test port
(PORT 1 for S
11
or PORT 2 for S
22
on ES analyzers).
NOTE Include any adapters that you will have in the device measurement. That is,
connect the standard device to the particular connector where you will
connect your device under test.
Figure 6-6 Standard Connections for a Response and Isolation Error
Correction for Reflection Measurements
8. To measure the standard, press:
or
If the calibration kit you selected has a choice between male and female calibration
standards, remember to select the sex that applies to the test port and not the standard.
CAL KIT [ ]
CAL KIT
SELECT CAL KIT
RETURN
CALIBRATE MENU
RESPONSE & ISOL’N
RESPONSE
SHORT
OPEN