Agilent Technologies Agilent E5250A Switch User Manual


 
Agilent E5250A User’s Guide, Edition 9 5-41
Controlling the E5250A
Using Virtual Front Panel Utility
Softkey Menu 2:
For the following conditions, the following softkeys are displayed by selecting the
MORE softkey on Softkey Menu 1.
Cursor points to row of normal input port or to row of input Bias Port with Bias
Mode OFF.
Softkey Menu 3:
For the following conditions, the following softkeys are displayed by selecting the
MORE softkey on Softkey Menu 1:
Cursor points to row of input Bias Port and Bias Mode is ON.
Softkey Description
OPEN
ALL
Opens all relays. If Bias Mode is ON, then closes relays to connect
bias enabled output ports to the input Bias Port.
OPEN
ROW
Opens all relays of row at cursor. If Bias Mode is ON, then closes
relays to connect bias enabled output ports (that are not connected
to other input ports) to the input Bias Port.
CLOSE
ROW
Closes all relays of row at cursor. And all output ports are
disconnected from the input Bias Port.
MORE Displays softkey menu 1.
RETURN Displays the previous level of softkey menu.
Softkey Description
OPEN
ALL
Opens all relays, then connects all bias enabled output ports to the
input Bias Port.
DISABLE
ROW
Bias disables all output ports. So relays open to disconnect all
output ports from the input Bias Port.
ENABLE
ROW
Bias enables all output ports, then connects all output ports (that
are not connected to another input port) to the input Bias Port.
MORE Displays softkey menu 1.
RETURN Displays the previous level of softkey menu.