Executing Self-Test and Leak Test
Using the Self-Test Utility
To execute the Leak Test, do the following:
1. On MAIN MENU of the E5250A Self-Test Utility, select LEAK TEST softkey.
The following message is displayed:
Input Agilent 4155/4156 GPIB address (default 817):?
2. To use default GPIB address 817 for the 4155/4156, just press Enter key.
If not, type GPIB address of the 4155/4156, then press
Enter key.
The LEAK TEST MENU with the following softkeys are displayed:
3. Select START TEST softkey. Following message/softkeys are displayed:
Select IV Port of Leak Test by softkey.
Softkey Description
START TEST For defining the Leak Test connection. Need to define the
E5250A input port and 4155/4156 SMU port.
MAIN MENU Returns to MAIN MENU of the Self-Test Utility.
Softkey Description
IVn (n=1 to 6) For specifying the E5250A SMU INPUTn port for which
you want to perform the leak measurement. If only the
E5255As are installed, only IVn softkeys displayed are for
SMU INPUTn internally connected to the E5255A IV input.
CANCEL Cancels the Leak Test execution.