Agilent Technologies Agilent E5250A Switch User Manual


 
10-4 Agilent E5250A User’s Guide, Edition 9
Specifications
Agilent E5252A 10x12 Matrix Switch
Agilent E5252A 10x12 Matrix Switch
Switch Specifications
These are specifications when the E5252A is installed in mainframe. In the table,
the conditions are as follows:
A: 23 °C ± 5 °C, 5 % to 60 % RH
B: 5 °C to 18 °C and 28 °C to 40 °C, 5 % to 60 % RH
C: 28 °C to 40 °C, 60 % to 80 % RH
A B C
Max. Current Rating (A) 1.0 1.0 1.0
Max. Voltage Rating (V)
Channel to Guard 200 200 200
Channel to Channel 300 300 300
Closed Channel Residual Resistance (Ω)
Low Leakage I-V Port 0.6 0.6 0.6
General I-V Port 1.0 1.0 1.0
C-V, HF Port 1.0 1.0 1.0
Channel Isolation Resistance (Ω)
Low Leakage I-V Port
10
13
5 × 10
12
2 × 10
12
General I-V Port
10
12
5 × 10
11
2 × 10
11
C-V, HF Port
10
9
5 × 10
8
2 × 10
8