Agilent Technologies Agilent E5250A Switch User Manual


 
Agilent E5250A User’s Guide, Edition 9 11-13
Error Messages
E5250A Specific Error Messages
3013 Cannot connect multiple channels in SROUTe mode
Incorrect channel list for Single connection rule. For Single
connection rule, an input port cannot be connected to multiple output
ports, and an output port cannot be connected to multiple input ports.
Change channel list or change connection rule to Free.
3014 Cannot directly specify Bias Port channel
Incorrect channel number when Bias Mode ON is specified in the
channel list. Cannot directly specify the channel number for the Bias
Port.
3015 Bad channel number combination on E5252A card
Wrong input port combination is specified in the channel list for the
E5252A card. For input ports, can specify only one of input port
numbers 5, 7, and 9, and only one of input port numbers 6, 8, and 10.
For example, channel list cannot specify input ports 5 and 7.
3016 Config error. Re-install E5255A.
May be problem with the E5255A installation. Re-install the
E5255A.
3017 Too many relays closed. Max 200.
Too many physical relays have been closed. To close new relays,
must open some relays. Maximum 200 relays can be closed.
3018 Can't change to ACONfig mode. Check card config.
The present card configuration of the E5250A does not allow the
Auto Config mode. Change the E5250A card configuration to allow
Auto configuration mode.
3019 Cannot use same port for Couple and Bias
Same input port number is specified to be Couple Port and Bias Port.
Couple Port and Bias Port cannot be the same port.