Agilent Technologies Agilent E5250A Switch User Manual


 
Agilent E5250A User’s Guide, Edition 9 3-19
Executing Self-Test and Leak Test
Using the Self-Test Utility
8. Connect the cable from the 4155/4156 SMU port specified by the SMUm
softkey (m=1 to 6) in step 5 to the E5250A input port specified by the IVn
softkey (n=1 to 6) in step 4.
Use Kelvin triaxial cable for Kelvin connection
Use triaxial cable for non-Kelvin connection
9. Press the CONTINUE softkey to execute the Leak Test.
After the Leak Test is completed for the input port specified by the IVn softkey,
the test result (PASS or FAIL) is displayed on the 4155/4156 screen. And the
following softkeys are displayed:
To display the leak current measurement data on the 4155/4156 screen, press the
REPORT CRT softkey.
10. To return to the LEAK TEST MENU, select the LEAK TEST softkey. And to
perform the Leak Test for another input port, repeat steps 3 to 9.
Softkey Description
REPORT CRT Displays the Leak Test results on the 4155/4156 screen.
LEAK TEST Returns to the LEAK TEST MENU.