Agilent Technologies Agilent E5250A Switch User Manual


 
Agilent E5250A User’s Guide, Edition 9 3-9
Executing Self-Test and Leak Test
Executing Self-Test
430 ENTER @Hp5250;A
440 IF A=0 THEN GOTO 470
450 DISP "Card 3 failed. Press Continue to test Card 4"
460 GOTO 480
470 DISP "Card 3 passed. Press Continue to test Card 4"
480 PAUSE
490 OUTPUT @Hp5250;":DIAG:TEST:CARD:EXEC? 4"
500 ENTER @Hp5250;A
510 IF A=0 THEN GOTO 540
520 DISP "Card 4 failed."
530 GOTO 550
540 DISP "Card 4 passed."
550 END
Line Number Description
30 to 90 Clears all previous test results.
100 to 170 Executes the Controller Test, then returns test result.
180 to 260 Executes the Front Panel Interface Test, then returns the test
result.
270 to 340 Executes the Relay Test for the card installed in slot 1 of the
E5250A, then returns test result.
350 to 410 Executes the Relay Test for the card installed in slot 2 of the
E5250A, then returns test result.
420 to 480 Executes the Relay Test for the card installed in slot 3 of the
E5250A, then returns test result.
490 to 550 Executes the Relay Test for the card installed in slot 4 of the
E5250A, then returns test result.