*ESE
The following table shows the bits of the Standard Event Status Register and
binary-weighted decimal value of each bit.
Example The following four lines enable the same bit (CME bit):
OUTPUT @Hp5250;"*ESE 32"
using decimal numeric
OUTPUT @Hp5250;"*ESE #B100000" using binary numeric
OUTPUT @Hp5250;"*ESE #Q40" using octal numeric
OUTPUT @Hp5250;"*ESE #H20" using hexadecimal numeric
The following is example for query:
OUTPUT @Hp5250;"*ESE?"
ENTER @Hp5250;A
bit
binary-
weight
description
0 1 OPC (Operation Complete)
1 2 not used
2 4 QYE (Query ERROR)
3 8 DDE (Device-Dependent ERROR)
4 16 EXE (Execution ERROR)
5 32 CME (Command ERROR)
6 64 not used
7 128 PON (Power on)