Agilent Technologies Agilent E5250A Switch User Manual


 
Agilent E5250A User’s Guide, Edition 9 4-19
Setting up Measurement Environment
Connecting the E5250A Input
Figure 4-7 Connection of Triax-Dual Triax Adapter
By this connection, input signal for SMU INPUT connectors is as follows:
If you make non-Kelvin connection for the E5250A outputs, use the odd input port
only. Because non-Kelvin connection uses only Force signal for measurement.
SMU INPUT AUX
To
SMU
1
2
3
4
5
6
Agilent E5250A
From
GND
TRIAX-DUAL TRIAX
Adapter
Agilent 16493H Cable
To
SMU INPUT
To
GNDU
SMU INPUT Port Number Input Signal
1, 3, or 5 Force
2, 4, or 6 Sense