Agilent Technologies Agilent E5250A Switch User Manual


 
9-34 Agilent E5250A User’s Guide, Edition 9
Executing Sample Programs
HCI Measurement Program
NOTE If you don't want program to pause, change the parameter in SP_MUX_A program to
Pause_to_save=0 before you run the SP_MUX_A program. The program
automatically extracts Tdc, saves result data, and stops.
5. All calculated data is saved to ANAHCI file on computer's disk drive, which is
an ASCII file. The data is also listed on the 4155/4156's screen. After analyzing,
saving, and listing the data, the 4155/4156 is initialized.
For example, in the case that you selected Idlin softkey in step 3, the following is
saved to the file:
Number of devices
Vdstr, Vgstr, Gate length, Gate width
Idlin
Device ,Validity ,Tdc_idlin
First device number, 0 or 1, Extracted Tdc for the device
:: :
Last device number, 0 or 1, Extracted Tdc for the device
Averaged Tdc_idlin
Calculated average Tdc
For the above, 0 means valid and 1 means invalid.
Following is an actual example for this case:
4
5, 1.95, 1.E-6, 1.E-5
Idlin
Device Validity Tdc_idlin
1, 0, 835.5786
2, 0, 3401.432
3, 0, 6269.047
4, 0, 24366.79
Averaged Tdc_idlin
8718.2119