Agilent Technologies Agilent E5250A Switch User Manual


 
Agilent E5250A User’s Guide, Edition 9 3-5
Executing Self-Test and Leak Test
Executing Self-Test
NOTE If Fail LED is ON after Relay Test
After executing the relay test, if the Fail LED turns on, the cause is one of the
following:
The Relay Test Adapter is not connected or a cable is still connected to input
connector.
Output is not open. See step 2 of previous procedure.
You did not properly install the E5255A. See “To Configure E5255A” on page
2-10.
A plug-in card may be defective. Contact your nearest Agilent Technologies
Service Center.
Figure 3-1 Relay Test Adapter Connection
NOTE Settings after Relay Test
After Relay Test, the E5250A settings are changed to same as after :SYST:CPON
command execution.
Relay Test Adapter