Agilent Technologies Agilent E5250A Switch User Manual


 
Agilent E5250A User’s Guide, Edition 9 3-7
Executing Self-Test and Leak Test
Executing Self-Test
Table 3-2 SCPI Commands for Self-Test
NOTE Front Panel Interface Test
After the :DIAG:TEST:FRAM:EXEC? FPAN command is executed, the Remote,
System Fail, Fail, and Local/Self Test LEDs on the E5250A front panel blink for
about 10 seconds. If an LED does not blink, the LED is defective.
Confirm that the four LEDs are blinking, then press the
Local/Self Test key on the
E5250A front panel within 10 seconds. The LEDs should stop blinking.
If you do not press the
Local/Self Test key within 10 seconds, the E5250A returns a
fail result for the Front Panel Interface Test, which turns on the Fail LED.
If you pressed
Local/Self Test key within 10 seconds, and the Fail LED still turns on,
it means the
Local/Self Test key is defective.
If you find any problem, contact your nearest Agilent Technologies Service Center.
Test Item Description SCPI Command
Controller
Test
Executes Controller Test,
then returns test result.
:DIAG:TEST:FRAM:EXEC? CONT
Returns test result. :DIAG:TEST:FRAM:STAT? CONT
Clears test result. :DIAG:TEST:FRAM:CLE CONT
Front Panel
Interface
(I/F) Test
Executes I/F Test, then
returns test result.
:DIAG:TEST:FRAM:EXEC? FPAN
Returns test result. :DIAG:TEST:FRAM:STAT? FPAN
Clears test result. :DIAG:TEST:FRAM:CLE FPAN
Relay Test Executes Relay Test, then
returns test result.
:DIAG:TEST:CARD:EXEC? ALL
or Card No. (1, 2, 3, or 4)
Returns test result. :DIAG:TEST:CARD:STAT? ALL
or Card No. (1, 2, 3, or 4)
Clears test result. :DIAG:TEST:CARD:CLE ALL or
Card No. (1, 2, 3, or 4)