Chapter 4, Making Measurements
Origin Offset
Origin Offset
Origin offset is measured in either Connect state or test mode. The measurement
procedure described below is for the Connect state. To measure in the test mode,
see “Test mode measurement procedure” on page 118.
Origin offset measures the magnitude of the RF carrier leakage relative to the
magnitude of the modulated carrier at the detection point.
Origin Offset measurement procedure
1. Make sure the Test Set is in Connect state. See
“Registration” on page 93.
2. Select
Meas
from the
Data/Meas
field.
3. Select one of the two measurement fields that can be modified.
4. Select
Orgin Ofs
from the list of choices. The Origin Offset measurement will be
displayed.
Figure 46 Origin Offset Display
Con-00dp.tif