Chapter 7, Screen and Field Descriptions
Parameters of the DIGITAL MEASUREMENT Screen
Amplitude Settings: 137.0 to +18.0
Units: dBm.
This is the amplitude of the RF signal that the Test Set will transmit. Note: you
may have to adjust the RF Gen Amplitude, Output Port, or Atten Hold when using
values above 0 dBm. Refer to the HP 8920 Series RF Test Set User Guide for
more information.
Note that the available setting is greater than both the specified maximum output
power of the Test System and the system specification as well. This is done to
allow you to use the maximum output power of each individual Test Set.
Dig Meas Settings: EVM1, EVM10, Adj Ch Pwr, Avg Pwr, BER, DTC WER, DCCH WER
Units: None.
This field shows the set of measurements that are being displayed.
DVCC Settings: 1 through 255
Units: None, Integer values.
This is the Digital Verification Color Code. This is for the DTC only. For the
DVCC of the DCCH, see “Parameters of the DCCH CALL CONFIGURE
Screen” on page 220.
Pwr Gain Settings: Auto, Hold
Units: dBm, Integer values.
Pwr Gain controls the RF attenuator and the DSP gain. Used with Average Power,
it is frequently best to keep this in Auto mode.