HP (Hewlett-Packard) HP 83206A Network Card User Manual


 
75
Chapter 3, Processing Calls
Authentication
Processing Calls
Perform an SSD Update
There are two methods that can be used to perform an SSD Update:
SSD Update on the Control Channel
SSD Update on the Voice Channel
Performing an SSD Update on the Control Channel
1. Ensure that the Test Set has Initialized Call Processing with Authentication. See
“Initialize Call Processing with Authentication through Registration” on page
72
for information on how to register the mobile station and activate Authentication.
2. Select AUTHEN on the To Screen meun.The AUTHENTICATION screen will be
displayed.
3. Select RANDSSD field. Enter a new, valid RANDSSD value (14-digit hexadecimal).
4. Select CALL CNTL under the To Screen menu. The CALL CONTROL screen will be
displayed.
5. Select SSD Upd on the CC Order field. The Access annunciator lights while the SSD
Update order is in progress. If the mobile station responds properly on the reverse
control channel, “SSD Update Order Successful” is displayed in the upper portion of
the CALL CONTROL screen. The Test Set and the mobile station are now linked and
further call processing functions with authentication can be tested. If the order failed,
then “SSD Update Order Failed” appears in the upper portion of the CALL
CONTROL screen.