Anritsu ML2430A Power Supply User Manual


 
Remarks:
Places a single ASCII character '1' on the GPIB output queue when the
conditions for the *OPC command are met. An operation is complete when
all input messages before the command have been completed and any
responses have been read out of the output buffer.
Example:
RGH A, 1; RGH B, 2; *OPC?
Returns a ‘1’ on the GPIB output when it has finished setting the range
hold commands.
*RCL Recall a stored setup
SYSTEM
Syntax:
*RCL <val>
val:
1 to 10
Remarks:
The ML2430A Series can store up to 10 instrument configurations for con
-
venient recall. The configuration parameters stored are:
Sensor Setup, Channel Setup, and Trigger Setup.
This command sets the ML2430A Series to a configuration previously
stored in memory locations 1 through 10. Trying to recall a setup that has
not been saved will set the execution bit in the event register (EXE in
ESR).
Related
Commands:
*SAV
*RST Reset Device
GPIB 488.2
Syntax:
*RST
Remarks:
Resets the ML2430A Series to the default configuration (see Appendix A,
Section A-3). Offset tables are not cleared. The GPIB address and emula
-
tion settings are not changed, and the input queue, output queue, and
status registers on the GPIB are not cleared. This command produces the
same result as the front panel key sequence System|Setup|PRE
-
SET|RESET.
*SAV Save configuration
SYSTEM
Syntax:
*SAV <val>
val:
1 to 10
Remarks:
Saves the configuration of the power meter into the memory location
specified. Sensor Setup, Channel Setup, and Trigger Setup are saved
along with all other instrument parameters.
ML2430A OM 6-15
GPIB OPERATION ML24XXA NATIVE COMMANDS