Anritsu ML2430A Power Supply User Manual


 
Remarks:
Selects the cal factor table to be used.
SM status message
Syntax:
SM
Remarks:
Returns the status message in the format:
AAaaBBCCccDDddEFGHIJKLMNOP<cr><lf>
where:
AA: measurement error code
aa: entry error code
BB: operating mode
CC: sensor A range
cc: 0
DD: sensor A filter
dd: 0
E: linear/log units
F: A
G: pwr ref status
H: REL mode status
I: trigger mode
J: group trigger mode
K: limits checking status
L: sensor A limits status
M: 0
N: offset status
O: duty cycle status
P: measurement units
SN Cal table identity update
Syntax:
SN<val>
val:
up to seven characters
Remarks:
Since the ML2430A stores the cal factor table information in the sensor,
this data must be saved to the sensor by using the ‘EX’ command or an
additional command ‘SV’, or the data could be lost. The saving of the cal
factor table data to the sensor can be done at the end of all updates to a
particular table.
ST Store setup
Syntax:
ST<val>EN
val:
1 to 10
6-114 ML2430A OM
HP 437B EMULATION COMMANDS GPIB OPERATION