Anritsu ML2430A Power Supply User Manual


 
Using
Buffered
Requests
Using the buffered Output channel ON command (page 6-64), even faster meas
-
urement speeds can be achieved. By using the ON command instead of the O
command
x
number of times, extra processing is removed, resulting in improved
speed.
NOTE
Using FAST mode here will not increase the speed as this
mode only works when asking for one measurement at a
time (i.e., the ‘O’ command only.)
C code example:
/* Reset unit and ask for 200 buffered readings n channel 1 */
Send(0, 13, “*RST; ON 1, 200”, 15L, NLend);
/* use a large buffer size (4K for 200 readings) */
Receive(0, 13, buffer, 4096, STOPend);
Settling (%) Power Level (dBm) Readings/Second
0.1 0 216
0.1 –30 202
Changing
measurement modes
By changing the sensor measurement mode to Custom, the ML2430A Series can
be precisely configured to meet the needs of the specific application.
In this example, the Trigger Gate Width (page 6-79) has been reduced to 1 ms,
Channel 1 is set to Sensor A, and Channel 2 is off.
C code example:
/* reset unit. Custom measurement mode, 1 ms TRGGW */
/* Auto averaging Sensor A Channel 2 off*/
/* Use FAST mode*/
Send(0, 13, “*RST; FAST ON; SENMM A,
CUSTOM; TRGGW 1ms”, 51L, NLend);
Receivesetup(0,13);
/* Fast mode, therefore do not send ‘O 1’ for data, just read.*/
for(i=0; i<10; i++)
RcvRespMsg(0,buffer,STOPend);
Power Level (dBm) Readings/Second
0 240
–30 240
ML2430A OM 5-23
PROCEDURES OPTIMIZING READINGS