Anritsu ML2430A Power Supply User Manual


 
Remarks:
Stores the present configuration to the selected register.
Example:
To store the current instrument configuration in register 2:
ST2EN
Related
Commands:
RC
SV Save cal factor table
Syntax:
SV
Remarks:
Since the ML2430A stores the cal factor tables in the sensors, this
command forces the edits to a cal factor table to be saved to the sensor.
The operation can take a couple of seconds to complete.
TR0 Trigger hold mode
Syntax:
TR0
Remarks:
Sets both channels to trigger hold mode. It does not trigger until it receives
a TR1 or TR2 or GET (group executive trigger), *TRG or TR3 command. If
it receives a TR3 command it reverts back to the trigger mode it was in be-
fore the TR0 command was sent. If the REM line is low, this command has
no effect.
Related
Commands:
TR1, TR2, TR3, *TRG, Group Execute Trigger (GET), GT0, GT1, GT2
TR1 Trigger immediate
Syntax:
TR1
Remarks:
Triggers a single reading which is added to the internal digital filter and the
updated filter power level is returned on the GPIB. After a TR1 command,
the instrument returns to TR0 standby mode.
Related
Commands:
TR0, TR2, TR3, *TRG, Group Execute Trigger (GET), GT0, GT1, GT2
TR2 Trigger with a settling delay
Syntax:
TR2
ML2430A OM 6-115
GPIB OPERATION HP 437B EMULATION COMMANDS