Anritsu ML2430A Power Supply User Manual


 
dB (rel)
Remarks:
Set to dB units in relative mode using the present relative reference value.
C dB (ref)
Remarks:
Set to dB units in relative mode using the present relative reference value,
and enable the application of the calfactor.
D dBm
Remarks:
Set units to dBm.
EMUL Select emulation mode
Syntax:
EMUL <mode>
mode:
ML24XX (Anritsu ML2430A Series native mode)
HP436A (Hewlett-Packard)
HP437B (Hewlett-Packard)
HP438A (Hewlett-Packard)
ML4803 (Anritsu ML4803A Series)
Remarks:
Sets the GPIB emulation to emulate other types of power meters. This
command is available in any emulation mode, and resets the whole GPIB
interface when the emulation mode is changed.
When selecting GPIB emulation modes, the instrument configures itself to
the preset conditions of the instrument to be emulated. For example,
when selecting HP 438A emulation, the front panel menus pass through
the presets for the HP 437B (which presets sensor A to dBm) then selects
HP 438A emulation (which presets sensor A to Watts).
NOTE
This command requires a white space between the com
-
mand header (EMUL) and the parameter <mode>. This is
an ML2430A-specific command that does not conform to
the HP Emulation command format defined at the begin
-
ning of this section.
H Hold mode
Remarks:
Sets both channels to trigger hold mode. The power meter does not trigger
until it receives an I or T command. If it receives an R or V command, it re
-
verts back to the trigger mode it was in before the H command was sent.
6-96 ML2430A OM
HP 436A EMULATION COMMANDS GPIB OPERATION