Anritsu ML2430A Power Supply User Manual


 
Syntax:
ON <c>,<val>
c:
1 or 2
val:
1 to 1000
Remarks:
Readout and Power vs. Time modes only. This command returns the speci
-
fied number of readings for the specified channel. The readings are first
assembled, and then passed to the GPIB as a whole, with a line feed char
-
acter (hex 0x0a) marking the end of the string.
Example:
ON 1, 9
This example will return:
–10.234, –10.234, –10.235, –10.238, -10.250, –10.270, –10.500, –10.934,
–12.234<0x0a>
OPMD ML2430A Series operation mode
SYSTEM
Syntax:
OPMD <mode>
mode:
DIGIT
PROFILE
PWRTIM
SRCSWP
Remarks:
This command selects the ML2430A Series operation mode (data collec-
tion method).
DIGIT = digital read out of channel data
PROFILE = profile of graphic display
PWRTIM = graph of channel power versus time
SRCSWP = source sweep graphic display
To use Graph output commands, the ML2430A Series must be in Profile or
Power vs. Time mode. To use the output channel data commands, the
ML2430A Series must be in the digital readout mode.
Related
Commands:
GRMD
Query:
OPMD?
Returned
String:
OPMD <mode>
PRINT Send details to the connected printer.
SYSTEM
Syntax:
PRINT
Remarks:
The type of printout depends on the operation mode currently selected. In
all modes, the printout includes a header with the current sensor settings
ML2430A OM 6-65
GPIB OPERATION ML24XXA NATIVE COMMANDS