Anritsu ML2430A Power Supply User Manual


 
This message can be manipulated to program a different table using the
CFULD command.
CFUSAV Cal factor table save
SENSOR
Syntax:
CFUSAV
Remarks:
This command saves the cal factor table currently being edited to the ap
-
propriate sensor. Processing may take a couple of seconds. Any command
that can select a new sensor and/or cal factor table for changing, will not
automatically save any previous changes made. It is the users responsibil
-
ity to issue a CFUSAV command.
CFUSEL Select cal factor table
SENSOR
Syntax:
CFUSEL <s>,<table number>
s:
A or B
table
number:
table number or combination to use
0 = factory default table
1 to 10 = user table being used
11 to 20 = factory table + user table being used
Remarks:
Selects the cal factor table or combination of tables to be used and auto-
matically updates the sensor.
Example:
CFUSEL A,13
Selects the factory table plus user table 3 in sensor A.
CFUTBL Number of cal factor tables in the sensor
SENSOR
Syntax:
CFUTBL <s>
s:
A or B
Remarks:
Returns the number of cal factor tables available in the selected sensor.
CFUUSE Number of cal factor table being used
SENSOR
Syntax:
CFUUSE <s>
s:
A or B
6-26 ML2430A OM
ML24XXA NATIVE COMMANDS GPIB OPERATION