c:
1, 2 or 1&2
val:
0.0 to 1.0 seconds
Remarks:
The time the system waits after a trigger event has happened
before taking measurements when in READOUT or POWER vs. TIME
mode.
Select channel 1, 2 or 1&2. Selecting 1&2 allows both channels to trigger
together on the same conditions without having to set up two sets of trig
-
ger data.
Query:
TRGDLY? <c>
Returned
String:
TRGDLY <c>,<val>
The TRG type commands return the trigger state of the selected channel if
the channel is ON. This depends on the settings of the “link triggers” flag
and the current mode of the sensor on the selected channel. If the channel
is OFF, the stored trigger state of the channel is returned.
TRGGW Set trigger gate width
TRIGGER
Syntax:
TRGGW <c>,<val>[units]
c:
1, 2 or 1&2
val:
100 ns to 7.0 seconds
Remarks:
The length of time the system uses to collect data when in READOUT or
POWER vs. TIME mode. The default value is 20 ms.
Select channel 1, 2 or 1&2. Selecting 1&2 allows both channels to trigger
together on the same conditions without having to set up two sets of trig
-
ger data.
Query:
TRGGW? <c>
Returned
String:
TRGGW <c>,<val>
The TRG type commands return the trigger state of the selected channel if
the channel is ON. This depends on the settings of the “link triggers” flag
and the current mode of the sensor on the selected channel. If the channel
is OFF, the stored trigger state of the channel is returned.
TRGLVL Set trigger level
TRIGGER
Syntax:
TRGLVL <c>,<val>
c:
1, 2 or 1&2
val:
–30 to +20 dBm
ML2430A OM 6-79
GPIB OPERATION ML24XXA NATIVE COMMANDS