Anritsu ML2430A Power Supply User Manual


 
2 cursor 2 reading
D P Power difference between cursor 1 and cursor 2
X1 X axis at cursor 1
X2 X axis at cursor 2
Data hold = MIN (or MAX) :
1 cursor 1 minimum reading, or maximum if MAX mode
2 cursor 2 minimum reading, or maximum if MAX mode
D P Power diff between cursor 1 and cursor 2 minimums
(or maximums if MAX mode)
X1 X axis at cursor 1
X2 X axis at cursor 2
Data hold = MIN&MAX:
1 cursor 1 MIN reading
cursor 1 MAX reading
2 cursor 2 MIN reading
cursor 2 MAX reading
CLEAR
[GPRST]
Available in Profile, Source Sweep and Power vs. Time modes
when the DATA HOLD representation selection is not set to
NORMAL (or AVERAGE for Power vs. Time). Pressing the CLEAR
key restarts the min/max collection.
LINK
CURSOR
[CURLK]
Links the screen cursors in Profile and Power vs. Time modes so
that when one is moved, both are moved. When the cursors are
linked, a horizontal bar is drawn between them on the screen. If
one cursor is moved, the other cursor moves with it to maintain
their relative positions and time interval between them. When the
cursors are linked, the relative time positions are altered by adjust
-
ing the gate width in the TRIGGER|SETUP|WIDTH submenu.
HOLD
[HOLD]
In Profile, Power vs. Time, and Source Sweep modes the graph
HOLD function allows a graph to be held and printed. The key
action is a toggle action, with the warning message Graph Display
HELD displayed at the top of the screen when HOLD is active.
Whenever measurement setup parameters are changed, graph
hold will automatically be released.
AUTO scale
[GRAUTO]
Auto scale for all graphic modes (Profile, Source Sweep and Power
vs. Time). In Profile and Source Sweep modes, auto scale will be
based on the min and max of the previous profile or sweep.
Display Controls the characteristics of the LCD display.
BACKLIGHT
[DBLGHT]
Controls the LCD backlight during internal battery operation. Can
be ON, OFF, or timed to go off after a specified period to save bat
-
tery life. The backlight is always on during AC or external DC power
operation.
ML2430A OM 4-27
OPERATION SYSTEM MENU