CCA Clear the calfactor value to zero.
CDJ Perform a CAL 0 dBm.
Remarks:
During the cal 0 dBm sequence, the CAL execution bit in the status byte is
set. When the CAL operation is completed, the CAL execution bit is
cleared.
COF Clear the offset value to zero.
COS Turn ON the 50 MHz, 0 dBm RF calibrator output.
CRF Clear the reference value to zero.
CST Turn OFF the 50 MHz, 0 dBm RF calibrator output.
DBM Sets the display channel units to dBm.
DBR Sets the display channel units to dB’s and takes the relative
value.
Remarks:
The relative value is stored as the reference data. The reference value can
be independently changed with the GPIB command REF.
EMUL GPIB emulation mode
Syntax:
EMUL <mode>
mode:
ML24XX (Anritsu ML2430A Series native mode)
HP436A (Hewlett-Packard)
HP437B (Hewlett-Packard)
HP438A (Hewlett-Packard)
ML4803 (Anritsu ML4803A Series)
Remarks:
Set the GPIB emulation to emulate other types of power meters. This com
-
mand is available in any emulation mode, and resets the whole GPIB inter
-
face when the emulation mode is changed.
6-86 ML2430A OM
ML4803A EMULATION COMMANDS GPIB OPERATION