MNGD <number_of_elements>,<element_1>,<element_2>,<ele
-
ment_n>...<\n>
The first number in the string is the number of elements to follow, and is al
-
ways 200 for the ML2430A Series.
MNMXS Track min and max values
CHANNEL
Syntax:
MNMXS <c>,<state>
c:
1 or 2
state:
ON or OFF
Remarks:
Turns ON or OFF the min/max tracking for the specified channel. The
MMRST command resets the values.
Related
Commands:
MMRST
Query:
MNMXS? <c>
Returned
String:
MNMXS <c>,<state>
MODDEL Modem redial delay time
SYSTEM
Syntax:
MODDEL <value>
value:
1 to 10
Remarks:
Sets the autodial delay between retrys. The value is the number of minutes
to delay between each autodial retry after a failure to connect. This interval
can be set from 1 to 10 minutes. See Section 5-10 for more information on
modem operation.
Query:
MODDEL?
Returned
String:
MODDEL <value>
MODINIT Initialize modem
SYSTEM
Syntax:
MODINIT
Remarks:
Initializes the modem connected to the ML2430A serial port. See Section
5-10 for more information on modem operation.
MODLIM Autodial enable for limits failure
SYSTEM
6-54 ML2430A OM
ML24XXA NATIVE COMMANDS GPIB OPERATION