Anritsu ML2430A Power Supply User Manual


 
Remarks:
Clears the cal factor table to one entry for 50 MHz at 100%, but does not
clear the identity of the table. The cleared table is automatically saved to
the sensor.
CFUID Cal factor table identity update
SENSOR
Syntax:
CFUID <s>,<table number>,<identity>
s:
A or B
table
number:
1 to number of tables supported by the sensor type
identity:
Seven characters or until a message terminator will be read as the identity.
Remarks:
Updates the seven character identity string. This only affects the copy of
the cal factor table stored in the memory of the power meter. To take affect
and not be lost, the table must be saved to the sensor using the CFUSAV
command.
Query:
CFUID? <s>,<table number>
Returned
String:
CFUID <s>,<table number>,<identity>
CFUNITS Cal factor display units
SENSOR
Syntax:
CFUNITS <s>,<units>
s:
A or B
units:
% or PCT
dB or dBm
Remarks:
This command changes the display units of the calfactors between either
dB or pecentage. Note that this will also set the form the data is output
over the GPIB (or RS232) when requested.
Query:
CFUNITS? <s>
Returned
String:
CFUNITS <s>,<units>
CFULD Cal factor table binary load
SENSOR
Syntax:
CFULD <s>,<table number>,<length>,<binary data>
s:
A or B
table
number:
1 to number of tables supported by the sensor type
length:
Length of message in bytes
6-24 ML2430A OM
ML24XXA NATIVE COMMANDS GPIB OPERATION