Anritsu ML2430A Power Supply User Manual


 
B–A, '5' = A/B, '6' = B/A, '7' = EXT Volts.
D = Channel 1 units: '0' = dBm, '1' = Watts, '2' = Volts, '3' = dBmV, '4' =
dBmV.
E = Channel 2 units: '0' = dBm, '1' = Watts, '2' = Volts, '3' = dBmV, '4' =
dBmV.
F = Channel 1 relative status: '0' = Rel OFF, '1' = Rel ON.
G = Channel 2 relative status: '0' = Rel OFF, '1' = Rel ON.
H = Channel 1 low limit state: '0' = OFF, '1' = ON.
I = Channel 1 high limit state: '0' = OFF, '1' = ON.
J = Channel 2 low limit state: '0' = OFF, '1' = ON.
K = Channel 2 high limit state: '0' = OFF, '1' = ON.
L = Sensor A measurement mode: '0' = Default, '1' = MOD average, '2' =
Custom.
M = Sensor B measurement mode: '0' = Default, '1' = MOD average, '2' =
Custom.
NN = Sensor A range hold: Manual = '01' to '05', AUTO = '11' to '15'.
OO = Sensor B range hold: Manual = '01' to '05', AUTO = '11' to '15'.
P = Sensor A averaging mode: ‘0’ = OFF, ‘1’ = AUTO, ‘2’ = Moving, ‘3’ =
Repeat.
Q = Sensor B averaging mode: ‘0’ = OFF, ‘1’ = AUTO, ‘2’ = Moving, ‘3’ =
Repeat.
RRRR = Sensor A average number. For Profile and Source Sweep
modes, this number is between 1 and 512. For digital Readout or Power
vs. Time modes, the values are either 1 to 512 or, if in AUTO averaging
mode, 513 to 1024.
SSSS = Sensor B average number (0000 if ML2437A). For Profile and
Source Sweep modes, this number is between 1 and 512. For digital
Readout or Power vs. Time modes, the values are either 1 to 512 or, if in
AUTO averaging mode, 513 to 1024.
T = Sensor A low level average: '0' = OFF, '1' = Low, '2' = Medium, '3' =
High.
U = Sensor B low level average: '0' = OFF, '1' = Low, '2' = Medium, '3' =
High.
ML2430A OM 6-73