Agilent Technologies B2201A Switch User Manual


 
4-12 Agilent B2200 User’s Guide, Edition 2
Programming
Programming Examples
Using Ground Mode
The following example uses the ground mode.
Figure 4-4 Ground Mode Example
Setup: Channel configuration mode: Auto
Connection rule: Single. Multiple connection is available for the ground port.
Connection sequence: Break_Before_Make
Used module: All switch modules installed in the mainframe.
Ground port: AUX Input 12
Ground enabled input port: SMU Input 5 to 8
Ground enabled output port: All output ports
Connection paths (connected when the ground mode is ON):
from ground port to all ground enabled input/output ports
Connection paths (connected by the :ROUT:CLOS command):
The following output ports will be disconnected from the ground port, and the
following input-output connections will be made.
from SMU1 to Output 1 (channel list 00101)
from SMU2 to Output 2 (channel list 00202)
from SMU3 to Output 3 (channel list 00303)
from SMU4 to Output 4 (channel list 00404)