Agilent Technologies B2201A Switch User Manual


 
7-14 Agilent B2200 User’s Guide, Edition 2
Error Messages
B2200 Specific Error Messages
3026 Cannot use same port for Unused and Couple
The couple mode and the ground mode cannot be used in parallel
when a couple port and the ground enabled input port (unused port)
have been assigned to the same input port.
3027 Cannot use Unused Port during Auto Ground Mode ON
The ground enabled input port (unused port) must be opened during
the ground mode is ON.
3030 Bad input port number
The input port number must be 1 to 14.
3031 Bad output port number
The output port number must be 1 to 12, 24, 36, or 48. Maximum
number depends on the switch module configuration.
3032 Bad setting memory number
The internal memory number for the setup data must be 1 to 8.
3033 EEPROM programming failure
Cannot change the GPIB address or update internal memory data.
Contact your nearest Agilent Technologies service center.
3034 EEPROM reading failure
Cannot read the GPIB address or internal memory data. Contact your
nearest Agilent Technologies service center.
3035 Setting memory data is invalid
Cannot read the setup data in the internal memory. The data has been
broken. Delete the data.
3036 Cannot load this setting data in this configuration
Cannot read the setup data in the internal memory due to the
configuration mismatch. Delete the data or leave it until restoring the
module configuration.
Error
Number
Error Message and Description