Agilent Technologies B2201A Switch User Manual


 
Agilent B2200 User’s Guide, Edition 2 2-7
Installation
Installing the B2200
Installing the B2200
This section describes how to install the Agilent B2200.
1. See “Requirements” on page 2-3, and determine the installation location.
2. Unpack the Agilent B2200 and place it at the installation site.
3. Plug the power cable into the Line input receptacle at the rear panel.
4. Plug the power cable into the power receptacle.
5. Perform the self-test. See “Self-Test” on page 2-10.
6. If you control the Agilent B2200 by using an external computer:
Set the GPIB address. See “To Set the GPIB Address” on page 2-8 and “To
Connect the GPIB Cable” on page 2-8.
7. If you have to add or replace the switch module:
Turn the Agilent B2200 off, remove the power cable, and install the switch
module. See “To Install the Switch Module” on page 2-9 and “To Install the
Blank Panel” on page 2-9.
NOTE Plug-in module is shipped from the factory after the specifications are confirmed.
Agilent can guarantee that the modules will function and that the module
performance is designed to meet its specifications. However, if you have any
trouble, contact your nearest Agilent Technologies service center.
If you require that the modules be guaranteed to their specifications (for ISO
compliance, etc.), then you must perform a calibration on the instrument (modules
and mainframe together). For calibration, contact your nearest Agilent Technologies
service center.