Agilent Technologies B2201A Switch User Manual


 
5-18 Agilent B2200 User’s Guide, Edition 2
SCPI Command Reference
Command Summary
DIAG subsystem :DIAGnostic subsystem has commands for executing the self-test function. For
more info, see “Selftest Menu” on page 3-34.
The :DIAGnostic subsystem commands ignore the B2200 channel configuration
mode. For :DIAG commands that require a card number, you specify 1, 2, 3, 4, or
ALL.
Command Description
:DIAG:TEST:CARD:CLE card_number Clears relay test result (pass/fail) of the specified card.
card_number: 1, 2, 3, 4, or ALL
:DIAG:TEST:CARD[:EXEC]? card_number Executes relay test, then returns result: 1 (fail card exists), 0
(pass).
card_number: 1, 2, 3, 4, or ALL
:DIAG:TEST:CARD:STAT? card_number Returns most recent relay test result: 1 (fail), 0 (pass), -1 (not
tested).
card_number: 1, 2, 3, 4
:DIAG:TEST:FRAM:CLE item Clears specified test result.
item: CONT (controller test), FPAN (front panel interface test),
LED, PEN, or BEEP
:DIAG:TEST:FRAM[:EXEC]? item Executes specified test, then returns test result: 1 (fail), 0
(pass).
item: CONT (controller test), FPAN (front panel interface test),
LED, PEN, or BEEP
:DIAG:TEST:FRAM:STAT? item Returns most recent test result of the specified test: 1 (fail), 0
(pass), -1 (not tested).
item: CONT (controller test), FPAN (front panel interface test),
LED, PEN, or BEEP