Agilent Technologies B2201A Switch User Manual


 
Agilent B2200 User’s Guide, Edition 2 7-11
Error Messages
B2200 Specific Error Messages
B2200 Channel Related Errors
Error
Number
Error Message and Description
2000 Invalid card number
Wrong card number is specified in card number or channel list
parameter. Make sure card is properly installed in the correct
B2200 slot.
2001 Invalid channel number
Wrong channel number is specified in the channel list. Confirm the
channel number, the card configuration, and the configuration
mode of the B2200.
2002 Unsupported Module
2003 Unsupported Configuration
2006 Command not supported on this card
Command was used that is not supported by the card.
2009 Too many channels in channel list
Too many channels are specified in the channel list (maximum
120).
2011 Empty channel list
No channel list is specified.
2012 Invalid channel range
Wrong channel list is defined. Confirm the syntax of the channel
list, the card configuration, and the configuration mode of the
B2200.