Agilent Technologies B2201A Switch User Manual


 
Agilent B2200 User’s Guide, Edition 2 2-3
Installation
Requirements
Requirements
This section describes the following requirements for the Agilent B2200.
“Power Requirements”
•“Power Cable
“Operating Environment”
“Storage and Shipping Environment”
Power Requirements
CAUTION Before applying ac line power to the Agilent B2200, ensure that the correct power
cable is used.
The Agilent B2200 can operate from any single-phase ac power source supplying 90
to 264 V in the frequency range from 47 to 63 Hz. The maximum power
consumption is 2 A/200 VA.
Power Cable
In accordance with international safety standards, this instrument is equipped with a
three-wire power cable. When connected to an appropriate ac power outlet, this
cable grounds the instrument frame. The type of power cable shipped with each
instrument depends on the country of destination. See the following table for the
part numbers of the power cables available.
If the plug on the cable does not fit the power outlet, or the cable is to be attached to
a terminal block, cut the cable at the plug end and re-wire it. This work should be
performed by a qualified electrician all local electrical codes being strictly observed.
The color coding used in the cable will depend on the cable supplied. If a new plug
is to be connected, it must meet local safety requirements and include the following
features:
Adequate load-carrying capacity.
Ground connection.
Cable clamp.