1-12 Agilent B2200 User’s Guide, Edition 2
Introduction
Specifications
Switch Modules
Table 1-1 lists the specifications when the switch module is installed in the
mainframe. In the table, the conditions are as follows:
Conditions: 23 °C ± 5 °C, 5 % to 60 % R.H.
Table 1-1 Agilent B2210A/B2211A Switch Module Specifications
The Agilent B2200A mainframe supports the Agilent B2210A module.
The Agilent B2201A mainframe supports the Agilent B2211A module.
Mixed configuration of the switch modules is not supported.
B2210A B2211A
Max. current rating (A)
SMU input 1.0 1.0
AUX input 0.5 0.5
Max. voltage rating (V)
SMU input, channel to common 200 200
SMU input, channel to channel 300 300
AUX input, channel to common 100 100
AUX input, channel to channel 100 100
Closed channel residual resistance ()
SMU input 0.6 0.6
AUX input 1.5 1.5
Channel isolation resistance ()
SMU input
10
14
5 × 10
13
AUX input
10
9
10
9