Agilent Technologies B2201A Switch User Manual


 
2-10 Agilent B2200 User’s Guide, Edition 2
Installation
Self-Test
Self-Test
NOTE To confirm the specifications
The self-test and diagnostics checks the operation of the mainframe and the
modules. However they cannot confirm if the Agilent B2200 satisfies its
specifications.
For verifying the specifications, contact your nearest Agilent Technologies Service
Center. Trained service personnel will perform calibration (performance
verification).
It is recommended to perform calibration once a year at least.
The following procedure performs the self-test and diagnostics.
1. Turn on the Agilent B2200.
The controller test will be performed. If the Agilent B2200 fails the test, contact
your nearest Agilent Technologies Service Center.
2. Press the
Shift key and the Menu key in this order. Self-test menu will appear on
the LCD.
SELF_TEST Controller test
RELAY_TEST Relay test
KEY Front panel interface test
BEEPER Beeper test
LED LED matrix test
PEN Light pen test
GPIB GPIB test
3. Move the cursor to the test item to perform, then press
Enter key.
4. To start the test, move the cursor to EXECUTE, then press
Enter key.
To display the previous test result, move the cursor to RESULT, then press
Enter
key.
For the details of each test, see “Selftest Menu” on page 3-34.