Agilent Technologies B2201A Switch User Manual


 
Agilent B2200 User’s Guide, Edition 2 7-5
Error Messages
Standard SCPI Error Messages
-120 Numeric data error
An error was detected in a numeric parameter (including the
non-decimal numeric types). This error message is reported when the
B2200 cannot determine the more specific errors -121 through
-128.
-121 Invalid character in number
An invalid character for the parameter was received; for example, an
alphacharacter was received when the parameter type was decimal
numeric.
-123 Exponent too large
The magnitude of the exponent for a numeric parameter was larger
than 32000.
-124 Too many digits
The mantissa of a decimal numeric parameter contained more than
255 digits excluding leading zeros.
-128 Numeric data not allowed
Numeric data is not allowed in this position for this command.
-138 Suffix not allowed
A suffix was received after a numeric parameter. For the B2200, no
parameters have suffix.
-140 Character data error
An error was detected in a character parameter. This error message is
reported if the B2200 cannot determine the more specific errors
-141 through -148.
-141 Invalid character data
Either the character parameter contains an invalid character or the
particular element received is not valid for the command.
-144 Character data too long
The character parameter contains more than 12 characters.
Error
Number