Agilent B2200 User’s Guide, Edition 2 7-13
Error Messages
B2200 Specific Error Messages
3017 Too many relays closed. Max 52 relays/card.
Too many relays have been closed. To close new relays, open some
relays. Maximum 52 relays can be closed for each module.
3018 Can't change to ACONfig mode. Check card config.
The present card configuration of the B2200 does not allow the auto
configuration mode. Change the module configuration.
3019 Cannot use same port for Couple and Bias
The couple mode and the bias mode cannot be used in parallel when
a couple port and the input bias port have been assigned to the same
input port.
3020 Bad auto ground port number
Specify the output ports effective for the input ground port. The port
number must be 1 to 14
3021 Bad unused port number
Specify the input ports effective for the ground enabled input port
(unused port). The port number must be 1 to 9.
3022 Cannot directly specify auto ground port channel
Relays on the input ground port cannot be controlled directly when
the ground mode is ON.
3023 Cannot directly specify unused port channel
Relays on the ground enabled input port (unused port) cannot be
controlled directly when the ground mode is ON.
3024 Cannot use same port for Couple and Auto Ground
The couple mode and the ground mode cannot be used in parallel
when a couple port and the input ground port have been assigned to
the same input port.
3025 Cannot use same port for Unused and Auto Ground
The input ground port and the ground enabled input port (unused
port) cannot be assigned to the same input port.
Error
Number
Error Message and Description