APPENDIX B SERVICE LOG ERROR MESSAGES
373XXA MM B-5
Error Message Description Corrective Action
ATTENUATOR UNAVAILABLE Option 6 Port 2 Test Step Attenuator
is not installed.
Install Option 6 Step Attenuator,
DIFFERENT H/W SETUP. RECALL
ABORTED
Model and/or options is (are) different
from the recalled setup.
Reconfigure system to duplicate the
hardware setup that was used to store
the saved data.
DIFFERENT S/W VERSION, RECALL
ABORTED
Saved state not compatible with soft
-
ware version or options.
Load compatible software (S/W) ver
-
sion and retry.
FREQUENCIES HAVE REACHED
UPPER LIMIT
Frequencies being defined in Multiple
Source mode have reached upper lim
-
its of Sources.
Redefine frequencies to not exceed
limits of Sources.
MEMORY LOCATION CORRUPTED Requested memory location is cor
-
rupted.
None. If problem reoccurs after stor
-
ing a new setup, contact ANRITSU
Customer Service.
NO BANDS ARE STORED No frequency bands have been de-
fined and stored.
Define and store frequency bands to
turn on Multiple Source mode.
NO STORED MEMORY DATA No data is stored in memory for dis-
play or trace math.
Store or re-save measurement data.
OPTION NOT INSTALLED Selected an option that is not in-
stalled.
None.
OUT OF CAL RANGE Entered values out of the selected
calibration range.
Change calibration range or re-enter
values that are within the current
range.
OUT OF H/W RANGE Entered value is out of the instru
-
ments hardware range.
Re-enter values that are within range.
OUT OF RANGE Entered value is out of range. Re-enter values that are within range.
RECEIVER OUT OF RANGE BY
EQUATION
Equation defined in Multiple Source
mode places receiver frequency out of
range when attempting to store band.
Redefine frequency.
SOURCE 1 OUT OF RANGE BY
EQUATION
Equation defined in Multiple Source
mode places Source 1 frequency out
of range when attempting to store
band.
Redefine frequency.
SOURCE 2 OUT OF RANGE BY
EQUATION
Equation defined in Multiple Source
mode places Source 2 frequency out
of range when attempting to store
band.
Redefine frequency.
Table B-1. Operational Error Messages (1 of 2)