Anritsu 373XXA Network Router User Manual


 
GPIB Interface Test This test verifies that the 373XXA printer interface is functioning cor
-
rectly.
Equipment Required:
GPIB cable (ANRITSU PN: 2100-2), or equivalent
CAUTION
Insure that no other cables are connected to either of the
two rear panel GPIB ports when performing this test.
Procedure:
Step 1. Press the Option Menu key (Enhancement key group) to dis
-
play the OPTIONS menu.
Step 2.
Select DIAGNOSTICS from menu; then in sequence se
-
lect: PERIPHERAL TESTS, and then GPIB INTER
-
FACE. See diagram below.
Step 3. Follow instructions on the screen until test is completed.
If test fails, proceed to Chapter 5, Troubleshooting.
OPERATIONAL TESTS PERIPHERALS AND INTERFACES
373XXA MM 3-9
OPTIONS
TRIGGERS
REAR PANEL
OUTPUT
DIAGNOSTICS
MULTIPLE SOURCE
CONTROL
PRESS <ENTER> TO
SELECT
DIAGNOSTICS
START SELF TEST
READ SERVICE LOG
INSTALLED OPTIONS
PERIPHERAL TESTS
TROUBLESHOOTING
(FOR SERVICE
ONLY)
HW CALIBRATIONS
(FOR SERVICE
ONLY)
PRESS <ENTER> TO
SELECT
PERIPHERAL TESTS
CRT DISPLAY
FRONT PANEL
EXTERNAL KEY-
BOARD
PRINTER INTERFACE
GPIB INTERFACE
PREVIOUS MENU
PRESS <ENTER> TO
SELECT