Anritsu 373XXA Network Router User Manual


 
SERVICE LOG ERROR MESSAGES APPENDIX B
B-6 373XXA MM
Error Message Description Corrective Action
STANDARD CAL NOT VALID FOR
WAVEGUIDE
Cannot use the standard method
when calibrating with waveguide.
Use the Offset Short method with
waveguide.
START F FOLLOWS PREVIOUS
STOP F
Start frequency of current band imme
-
diately follows stop frequency of previ
-
ous band. Cannot be modified.
None.
START MUST BE LESS THAN STOP Entered start frequency is greater
than the stop frequency.
Re-enter frequency values such that
the start frequency is lower than the
stop frequency.
STEP IS TOO LARGE Entered harmonic frequency extends
the stop out of range.
Re-enter so that harmonic frequency
is within range.
STOP IS OVER RANGE Entered value exceeds the instru
-
ments stop frequency.
Re-enter stop frequency.
SYSTEM NOT CALIBRATED 373XXA is uncalibrated for the se-
lected measurement values.
Perform a measurement calibration.
TOO FEW POINTS, 2 MINIMUM Entered too few discrete fill points, 2
is minimum.
Re-enter data points.
TOO MANY POINTS, 1601 MAXI-
MUM
Entered too many discrete fill points,
1601 points are the maximum al-
lowed.
Re-enter data points.
UNDEFINED DIVIDE BY ZERO Denominator cannot be zero in equa
-
tion.
Make denominator a value other than
zero.
WINDOW TOO SMALL Attempted to set time domain range
smaller than allowed
Re-enter larger time range.
OUT OF WINDOW RANGE Attempted to set time domain range
larger than allowed
Re-enter values within allowed range.