SERVICE LOG ERROR MESSAGES APPENDIX B
B-10 373XXA MM
Error Message Description
Invalid cal term for calibration
type
An attempt was made to program a calibration term which does not exist for the cur
-
rent calibration type.
Front panel setup not valid An attempt was made to get a front panel setup that did not contain a correct/valid
state.
Normalization data not valid An attempt was made to reference normalization data when there was no normaliza
-
tion data currently stored.
Command sequence too long An attempt was made to define a device trigger command sequence which had more
than 255 characters.
Unable to display menu An attempt was made to display a menu which could not be displayed for the current
373XXA state.
String too long An attempt was made to enter a string for the following mnemonics which exceeded
the specified maximum length.
LTD, LID, LMS and LNM - maximum length is 15 characters.
LOC - maximum length is 79 characters.
Must specify a calibration type
first
In order to perform a calibration, the calibration type must be specified by the use of
one of the Cxx mnemonics (i.e. C12, C8T, etc.) PRIOR to the issuance of the mne-
monics CWC, TDC or BEG.
Parameter value unchanged An attempt was made to change a start/stop frequency or number of data points to a
value outside of the current calibrated range with correction turned on.
Parameter change not permitted An attempt was made to perform an illegal state change or action based on the cur
-
rent 373XXA state. This includes attempting to store an undefined band definition. Or
certain changes from the calibration state or the calibration define state when defining
discrete frequencies.
Parameter value out of range
Parameter out of hardware
range
An attempt was made to set a parameter to a value outside of the permissible range
of values for the parameter.
Standard cal method not valid
for waveguide
In a waveguide type of calibration, the standard (OSL) cal method is forbidden.
Out of calibrated range An attempt was made to change a parameter not permitted to be changed with correc
-
tion on.
Start must be must be less than
stop
An attempt was made to set a new start frequency, distance or time greater than or
equal to the current stop frequency, distance or time. Or to set a new stop frequency,
distance or time less than or equal to the current start frequency, distance or time.
Tune mode requires a 12 term
calibration
Perform a 12 term calibration prior to turning on tune mode.
Table B-3. GPIB-Related Error Messages (3 of 8)