Anritsu 373XXA Network Router User Manual


 
quence: H/W CALIBRATIONS, and then SOURCE ALC
CALIBRATION (below).
Step 3. Follow the directions displayed on the screen to set-up and
connect the power meter to the 373XXA. Refer also to Fig
-
ure 6-2. Select START ALC CALIBRATION from menu.
Step 4. Follow the directions displayed on the 373XXA CRT screen,
until calibration is completed.
Step 5. If the software allows a Port 2 ALC Calibration, perform it
now.
Post Calibration
Actions
After the calibration process is completed, perform the following ac
-
tions, as appropriate:
RF POWER/ALC CALIBRATION ADJUSTMENTS
6-10 373XXA MM
789
45
6
12
3
0
.
-
SC6230 Sensor
ANRITSU 373XXA
VNA
Dedicated GPIB
ANRITSU ML24XXA
Power Meter
Figure 6-2. Equipment Set-Up for RF Power/ALC Calibration
OPTIONS
TRIGGERS
REAR PANEL
OUTPUT
DIAGNOSTICS
MULTIPLE SOURCE
CONTROL
PRESS <ENTER> TO
SELECT
DIAGNOSTICS
START SELF TEST
READ SERVICE LOG
INSTALLED OPTIONS
PERIPHERAL TESTS
TROUBLESHOOTING
(FOR SERVICE
ONLY)
H/W CALIBRATIONS
(FOR SERVICE
ONLY)
PRESS <ENTER> TO
SELECT
HARDWARE
CALIBRATIONS
(FOR SERVICE
ONLY)
LO1
CALIBRATION
LO2
CALIBRATION
SOURCE FREQUENCY
CALIBRATION
SOURCE ALC
CALIBRATION
DISK OPERATIONS
PREVIOUS MENU
PRESS <ENTER> TO
SELECT
HARDWARE
CALIBRATIONS
(FOR SERVICE
ONLY)
START SOURCE
FREQ CALIBRATION
PREVIOUS MENU
PRESS <ENTER> TO
SELECT
Set-Up