National Instruments AT-MIO-16X Switch User Manual


 
Chapter 4 Register Map and Descriptions
©
National Instruments Corporation 4-5 AT-MIO-16X User Manual
Command Register 1
Command Register 1 contains 11 bits that control AT-MIO-16X serial
device access, and data acquisition mode selection. The contents of this
register are not defined upon power up and are not cleared after a reset
condition. This register should be initialized through software.
Address: Base address + 00 (hex)
Type: Write-only
Word Size: 16-bit
Bit Map:
Bit Name Description
15 EEPROMCS EEPROM Chip Select—This bit controls
the chip select of the onboard EEPROM
used to store calibration constants. When
EEPROMCS is set, the chip select signal
to the EEPROM is enabled. Before
EEPROMCS is brought high, SCLK
should first be pulsed high to initialize
the EEPROM circuitry.
14 SDATA Serial Data—This bit isused to transmit a
single bit of data to the EEPROM and
both of the calibration DACs.
13 SCLK Serial Clock—A low-to-high transition
of this bit clocks data from SDATA into
the EEPROM (when EEPROMCS is set)
and the calibration DAC. If EEPROMCS
is cleared, toggling SCLK does not affect
the EEPROM. Serial data is always
loaded into the calibration DACs, but the
information is not updated until after the
76 5
SCN2
43
RTSITRIG
2
0
10
0SCANEN CNT32/16* 00
15
MSB
LSB
14 13
SDATA
12 11
SCANDIV
10
INTGATE
98
RETRIG_DISEEPROMCS SCLK DAQEN0