Chapter 4 Register Map and Descriptions
©
National Instruments Corporation 4-73 AT-MIO-16X User Manual
RTSI Switch Strobe Register
The RTSI Switch Strobe Register is written to in order to load the
contents of the RTSI Switch Shift Register into the RTSI Switch
Control Register, thereby updating the RTSI switch routing pattern. The
RTSI Switch Strobe Register is written to after shifting the 56-bit
routing pattern into the RTSI Switch Shift Register.
Address: Base address + 0E (hex)
Type: Write-only
Word Size: 8-bit
Bit Map: Not applicable, no bits used